Publications

Uniformity Correction of a Small Animal SPECT System with Spiral Step Data Acquisition

Abstract

1807
Objectives
A small animal (SA) SPECT system acquires data by stepping the scanning bed in a spiral pattern while translating in the axial direction. When using the fast data acquisition mode (FDAM), pitch is increased to shorten scan time. As a tradeoff, uniformity flood (UF) images have distinct non-uniformity patterns. Our aim is to develop and evaluate a uniformity correction (UC) method for images acquired with the FDAM.
Methods
We studied the SA SPECT system’s spiral step sequences with user defined scan volumes (UDSVs) of different sizes while using the FDAM. Each UDSV had a transaxial cross-section of 28 mm in diameter and different axial lengths (ALs). Sets of UDSVs with certain ALs were found to have UF images with matching non-uniformities. Using one UDSV set, 4 syringes were filled with 10, 15, 20, and 25 ml of ~0.4 mCi/ml 99mTc solution and imaged with the FDAM using ALs of 49, 63 …

Metadata

publication
Journal of Nuclear Medicine 56 (supplement 3), 1807-1807, 2015
year
2015
publication date
2015/5/1
authors
Andrew Rittenbach, Chao Wu, Frederik Beekman, Benjamin Tsui
link
https://jnm.snmjournals.org/content/56/supplement_3/1807.short
source
Journal of Nuclear Medicine
volume
56
issue
supplement 3
pages
1807-1807
publisher
Society of Nuclear Medicine