System calibration method for a CZT detector based ring-type small animal SPECT system
Abstract
2392
Objectives
Accurate system calibration (SC) is important for the performance of a CZT detector based SPECT system. Our goal was to develop a SC method for a ring-type SPECT system composed of 95 MR-compatible CZT detector modules for small animal SPECT-MR imaging.
Methods
The detector pixel uniformity was measured using a custom-made annular shell phantom with uniform shell thickness that fits tightly inside the ring SPECT system without the collimator. The phantom was filled with 15 µCi of Tc-99m solution and listmode data were acquired from the system. From the data, a 10% energy window was centered around the Tc-99m photopeak for each pixel and each window’s position on its pixel’s pulse height spectrum was stored for processing experimental data. The uniform flood image obtained with the energy calibrated data was used for uniformity correction of the detector. Pixels in the …